SPURIOUS COUNTS FROM STATIC PROBLEMS
SUGGESTIONS FOR A CLIENT IN ISRAEL

Static can and usually does cause spurious counts by activating the phosphor inside the cell. The solution usually is as simple as using a static spray on the PVC parts of the chamber. Purchase a good static spray at one of the chemical spray manufacturers. EN/Stat #MS-966 by Miller-Stephenson Chemecial Co. Inc, Danbury, CT, U.S.A. (Phone 203-743-4447) is an excellent static spray. You may also be able to find a static spray in at a TV supply store, however, I know that this one works and does not cause other problems, such as melting the foam or PVC.
  • 1) Remove the foam from the top of the chamber cap and spray both the foam and the inside and outside of the PVC chamber cap.
  • 2) Let the foam and the PVC chamber cap dry thoroughly and then reassemble.
  • 3) Don't forget to spray the chamber extension and its foam insert as well.
  • 4) This spray job will last you for a long time, you may only have to do this once or twice a year, however long it takes to wear off the coating, through handling.
  • 5) Make sure that the foam is not holding the chamber cap from completely seating on to the aluminum lower section.
  • 6) If the foam is keeping the PVC cap from seating you will let in some light and you might have to either adjust the foam thickness by re-seating or carefully trimming the foam.
  • 7) Please inspect your chamber and cap fit before you make modifications..
  • 8) Do not hesitate to E-mail me if you have further questions.